ES-2462 Rev.15

M5TS

Super-mini Terminal Block Signal Conditioners M5-UNIT

THERMOCOUPLE TRANSMITTER

Functions & Features
• Accepts direct input from a thermocouple and provides a standard process signal
• Linearization and burnout
• Cold junction compensation
• Fast response type available
• High-density mounting
• Power LED

MODEL: M5TS-[1][2]-[3][4]

ORDERING INFORMATION

• Code number: M5TS-[1][2]-[3][4]
Specify a code from below for each of [1] through [4].
  (e.g. M5TS-2A-R/K/BL/Q)
• Temperature range (e.g. 0 – 800°C)
• Special output range (For codes Z, 0 & 01)
• Specify the specification for option code /Q
 
(e.g. /C01/V01/S01)

[1] INPUT THERMOCOUPLE

1: (PR) (Usable Range 0 to 1760°C, 32 to 3200°F)
2: K (CA) (Usable range -270 to +1370°C, -454 to +2498°F)
3: E (CRC) (Usable range -270 to +1000°C, -454 to +1832°F)
4: J (IC) (Usable range -210 to +1200°C, -346 to +2192°F)
5: T (CC) (Usable range -270 to +400°C, -454 to +752°F)
6: B (RH) (Usable range 0 to 1820°C, 32 to 3308°F)
7: R (Usable range -50 to +1760°C, -58 to +3200°F)
8: S (Usable range -50 to +1760°C, -58 to +3200°F)
N: N (Usable range -270 to +1300°C, -454 to +2372°F)
0: Specify

[2] OUTPUT

Current
A: 4 – 20 mA DC (Load resistance 550 Ω max.)
D: 0 – 20 mA DC (Load resistance 550 Ω max.)
Z: Specify current (See OUTPUT SPECIFICATIONS)
Voltage
1: 0 – 10 mV DC (Load resistance 100 kΩ min.)
(CE or UKCA not available)
2: 0 – 100 mV DC (Load resistance 100 kΩ min.)
(CE or UKCA not available)
3: 0 – 1 V DC (Load resistance 100 Ω min.)
4: 0 – 10 V DC (Load resistance 1000 Ω min.)
5: 0 – 5 V DC (Load resistance 500 Ω min.)
6: 1 – 5 V DC (Load resistance 500 Ω min.)
1W: -10 – +10 mV DC (Load resistance 100 kΩ min.)
(CE or UKCA not available)
2W: -100 – +100 mV DC (Load resistance 100 kΩ min.)
(CE or UKCA not available)
3W: -1 – +1 V DC (Load resistance 800 Ω min.)
4W: -10 – +10 V DC (Load resistance 8000 Ω min.)
5W: -5 – +5 V DC (Load resistance 4000 Ω min.)
0: Specify voltage (See OUTPUT SPECIFICATIONS)
01: Specify voltage (See OUTPUT SPECIFICATIONS)
(CE or UKCA not available)

[3] POWER INPUT

AC Power
M: 85 – 264 V AC (Operational voltage range 85 – 264 V,
47 – 66 Hz)
(CE or UKCA not available)
DC Power
R: 24 V DC
(Operational voltage range 24 V ±10 %, ripple 10 %p-p max.)

[4] OPTIONS (multiple selections)

Response Time (0 – 90 %)
blank: Standard (≤ 0.5 sec.)
/K: Fast Response (Approx. 25 msec.)
Burnout
blank: Upscale burnout
/BL: Downscale burnout
/BN: No burnout
Other Options
blank: none
/Q: Option other than the above (specify the specification)

SPECIFICATIONS OF OPTION: Q (multiple selections)

COATING (For the detail, refer to our web site.)
/C01: Silicone coating
/C02: Polyurethane coating
/C03: Rubber coating
ADJUSTMENT
/V01: Multi-turn fine adjustment
TERMINAL SCREW MATERIAL
/S01: Stainless steel

GENERAL SPECIFICATIONS

Construction: Terminal block
Connection: M3.5 screw terminals (torque 0.8 N·m)
Screw terminal: Nickel-plated steel (standard) or stainless steel
Housing material: Flame-resistant resin (black)
Isolation: Input to output to power
Overrange output: Approx. -10 to +110 % at 1 – 5 V
Zero adjustment: -2 to +2 % (front)
Span adjustment: 98 to 102 % (front)
At burnout: Downscale ≤ -10 %, Upscale ≥ 110 %,
For output code 4W, Downscale ≤ -3%, Upscale ≥ 103%
Linearization: Standard
Cold junction compensation: CJC sensor attached to the
input terminals
Power indicator LED: Green LED turns on when the power is supplied.

INPUT SPECIFICATIONS

Minimum span: 3 mV
Input resistance: 20 kΩ min.
Burnout sensing: 0.1 μA

Lower temperature range & span (in °C)
(PR): Lower range 0 to 880°C;  min. span 370°C
K (CA): Lower range -270 to +1200°C;  min. span 75°C
E (CRC): Lower range -270 to +750°C;  min. span 50°C
J (IC): Lower range -210 to +800°C;  min. span 60°C
T (CC): Lower range -270 to +325°C;  min. span 75°C
B (RH): Lower range 0 to 750°C;  min. span 780°C
R: Lower range -50 to +550°C;  min. span 360°C
S: Lower range -50 to +550°C;  min. span 380°C
N: Lower range -270 to +1100°C;  min. span 110°C

Lower temperature range & span (in °F)
(PR)
: Lower range 32 to 1616°F;  min. span 670°F
K (CA): Lower range -454 to +2192°F;  min. span 140°F
E (CRC): Lower range -454 to +1382°F;  min. span 90°F
J (IC): Lower range -346 to +1472°F;  min. span 110°F
T (CC): Lower range -454 to +617°F;  min. span 140°F
B (RH): Lower range 32 to 1382°F;  min. span 1450°F
R: Lower range -58 to +1022°F;  min. span 680°F
S: Lower range -58 to +1022°F;  min. span 700°F
N: Lower range -454 to +2012°F;  min. span 200°F

Note 1: The described accuracy may be partially not satisfied when the temperature ranges below 0°C. Consult us.
Note 2: Consult us for the range out of the above list.

OUTPUT SPECIFICATIONS

■ DC Current: 0 – 20 mA DC
Minimum span: 1 mA
Offset: Max. 1.5 times span
Load resistance: Output drive 11 V max.
DC Voltage
Output code 0 (CE)
  Voltage range: -10 – +10 V DC
  Minimum span: 1 V
  Offset: Max. 1.5 times span
  Load resistance: Output drive 10 mA max.; at ≥1 V
  Max. 1.25 mA output drive for negative voltage
Output code 01 (Not CE)
  Voltage range: -1 – +1 V DC
  Minimum span: 10 mV
  Offset: Max. 1.5 times span
  Load resistance: Min. 100 kΩ
  Min. 1 MΩ for negative voltage

INSTALLATION

Power Consumption
AC:
  Approx. 2 VA at 100 V
  Approx. 3 VA at 200 V
  Approx. 3 VA at 264 V
DC: Approx. 2 W
Operating temperature: -5 to +55°C (23 to 131°F)
Operating humidity: 0 to 90 %RH (non-condensing)
Mounting: DIN rail
Weight: 80 g (2.8 oz)

PERFORMANCE in percentage of span

Accuracy (whichever is greater)
  K, E, J: ±0.1 % of FS or ±0.2°C (±0.36°F)
  T, N: ±0.2 % of FS or ±0.2°C (±0.36°F)
  PR, B, R, S: ±0.3 % of FS
    (at over 400°C for R, S and PR; over 770°C for B)
Cold junction compensation error
      (at 20°C ±10°C or 68°F ±18°F)
  K, E, J, T, N: ±0.5°C or ±0.9°F
  S, R, PR: ±1°C or ±1.8°F
Temp. coefficient: ±0.015 %/°C (±0.008 %/°F)
  (at over 400°C or 750°F for R, S and PR; over 770°C or 1420°F for B)
Burnout response: ≤ 10 sec.
Line voltage effect: ±0.1 % over voltage range
Insulation resistance: ≥ 100 MΩ with 500 V DC
Dielectric strength (input to output to power to ground)
  DC powered: 2000 V AC @1 minute
  AC powered: 1500 V AC @1 minute

STANDARDS & APPROVALS

EU conformity:
EMC Directive 
  EMI EN 61000-6-4
  EMS EN 61000-6-2
RoHS Directive
UK conformity (UKCA):
The UK legislations and designated standards are equivalent to the applicable EU directives.
(Refer to our website for more information about the legislations and designated standards.)

FRONT VIEW

EXTERNAL DIMENSIONS unit: mm [inch]

TERMINAL ASSIGNMENTS unit: mm [inch]

SCHEMATIC CIRCUITRY & CONNECTION DIAGRAM