ES-5984 Rev.3
M1EXR-2
2 Channel Super-mini Signal Conditioners
with Display M1E-2 Series
RTD TRANSMITTER
(2 channels, PC programmable)
Functions & Features
• Accepts direct input from a RTD and provides a linearized process signal
• Wide selection of RTD
• 2 channels
• Adjustments with the front buttons
• Leadwire resistance compensation
• OEL display with good visibility
Typical Applications
• Long distance transmission between the RTD and the transmitter
MODEL: M1EXR-2–[1][2]
ORDERING INFORMATION
• Code number: M1EXR-2-[1][2]
Specify a code from below for each of [1] and [2].
(e.g. M1EXR-2-M2/Q)
• Specify the specification for option code /Q
(e.g. /C01/SET)
NO. OF CHANNELS
2
: 2 channels
INPUT – Field-selectable
RTD
JPt 100 (JIS'89), Pt 100 (JIS'89), Pt 100 (JIS'97, IEC),
Pt 50Ω (JIS'81), Ni 508.4Ω, Pt 1000, Ni 100, Cu 10 @ 25℃
OUTPUT – Field-selectable
◆ DC Output
•
Current output
: 0 - 20 mA DC
• Voltage output
: -5 - +5 V DC
• Voltage output
: -10 - +10 V DC
[1] POWER INPUT
AC Power
M2
: 100 – 240 V AC (Operational voltage range 85 – 264 V,
47 – 66 Hz)
DC Power
R
: 24 V DC
(Operational voltage range 24 V ±10 %, ripple 10 %p-p max.)
P
: 110 V DC
(Operational voltage range 85 – 150 V, ripple 10 %p-p max.)
[2] OPTIONS
blank
: none
/Q
: With options (specify the specification)
SPECIFICATIONS OF OPTION: Q (multiple selections)
COATING (For the detail, refer to our web site.)
/C01
: Silicone coating
/C02
: Polyurethane coating
/C03
: Rubber coating
EX-FACTORY SETTING
/SET
: Preset according to the Ordering Information Sheet
(No. ESU-5984)
RELATED PRODUCTS
• Installation base (model: M1E-BS2)
• M1E configurator software (model: M1ECFG)
Downloadable at our web site.
A dedicated cable is required to connect the module to the PC. Please refer to the internet software download site or the users manual (EM-5981) for the M1E configurator for applicable cable types.
GENERAL SPECIFICATIONS
Construction
: Plug-in
Connection
• Connected to base with connector
• Base
M2.6 screw terminals (torque: 0.5 N·m)
Applicable solderless terminal size (M3)
Screw terminal
: Nickel-plated steel
Housing material
: Flame-resistant resin (black)
Isolation
: Ch1.input to ch2.input to Ch1.output to ch2.output to power
Overrange output
: Approx. -10 to +110 % (Not available for current output not greater than 0 mA)
Input zero fine adj. range
: -10 to +10℃ (front button)
Input span fine adj. range
: -10 to +10℃ (front button)
Output zero fine adj. range
: -5 to +5% (front button)
Output span fine adj. range
: 95 to 105% (front button)
Burnout protection
: Upscale standard; downscale or no burnout optional by programming
Configuration
: Program with front buttons or Via PC.
Programmable features include
: Refer to the parameter list.
Configurator connection
: 2.5 dia. miniature jack;
RS-232-C level
■Recommended solderless terminal size - M3 (unit: mm)
DISPLAY
Display functions
: Displays and sets measured range, engineering unit
Display size
: Approx. 15.6 × 20.8 mm (0.61” × 0.82”)
Number of pixels
: 68 × 95 (horizontal × vertical)
Character color
: Yellow
Display life
: Approx. 50000 hours
(Expected time for the Display brightness to be reduced to 50 % when the Display is used continuously with brightness setting 2 in 25 ℃)
Display type
: OEL display
Display digit
: Negative 5 digits, positive 6 digits (-99999 to 999999)
Decimal point position
: None, number of decimal places 1 or 2
INPUT SPECIFICATIONS
Input
: 2- or 3-wire RTD
Maximum leadwire resistance
: 50 Ω per wire (For Cu 10, 20 Ω per wire)
Sensing current
:
Pt 100, JPt 100, Pt 50 Ω, Ni 100, Cu 10: approx. 1.0 mA
Pt 1000, Ni 508.4 Ω: approx. 0.2 mA
Measurement rang
: -5 to +105% of input setting value
Refer to the operating manual (EM-5984-B) for changing input range, setting value.
OUTPUT SPECIFICATIONS
■ DC Current
Output range
: 0 – 20 mA DC
Conformance range
: 0 – 22mA DC
Minimum span
: 1 mA
Load resistance
: Output drive 15 V max.
(e.g. 4 – 20 mA: 750 Ω [15 V ÷ 20 mA])
(Not available to output less than 0 mA, output range may not be extended to -5 %)
■
DC Voltage
Output range
: -5 - +5 V DC, -10 - +10 V DC
Conformance range
: -6 - +6 V DC, -12 - +12 V DC
Minimum span
: 250 mV, 1 V
Load resistance
: 200 kΩ min.
INSTALLATION
Power Consumption
•
AC
:
≤ 7 VA at 100 V
≤ 10 VA at 200 V
≤ 12 VA at 264 V
•
DC
: ≤ 4 W
Operating temperature
: -5 to +55°C (23 to 131°F)
Storage temperature
: -5 to +55°C (23 to 131°F)
Operating humidity
: 10 to 85 %RH (non-condensing)
Mounting
: Surface or DIN rail
Weight
: 120 g (0.26 lb) except base
PERFORMANCE
Accuracy
: Input accuracy + output accuracy
See CALCULATION EXAMPLES OF OVERALL ACCURACY.
Inversely proportional to the setting span.
Input accuracy
: (whichever is greater)
Table 1 accuracy value or ±0.1 % of FS
Display accuracy
:
Input display: Input accuracy ±1 digit
Output display: Input accuracy + output accuracy ±1 digit
Output accuracy
: (% of max. output range)
0 - 20 mA: ±0.06 %
-5 - +5 V: ±0.03 %
-10 - +10 V: ±0.03 %
Temp. coefficient (% of max. I/O range)
: ±0.015 %/°C (±0.008 %/°F)
Response time (filter time constant: 0 sec.)
: ≤ 0.5 sec.
(0 – 90 %)
Burnout response
: ≤ 10 sec.
Line voltage effect
: ±0.1 % over voltage range
Insulation resistance
: ≥ 100 MΩ with 500 V DC
Dielectric strength
: 1500 V AC @ 1 minute
(Ch1 input to Ch2 input to Ch1 output to Ch2 output to power to ground)
CALCULATION EXAMPLES OF OVERALL ACCURACY
[Example] Input Range Pt 100 (JIS'97, IEC), 0 – 100°C, Output Range 0 – 20 mA, Output Setting 4 – 20 mA
■
Overall accuracy
• Input Accuracy
100 ℃ × 0.1 % = 0.1 ℃
0.2 ℃ is higher thus 0.2 ℃ is selected
0.2 ℃ ÷ 100 ℃ = 0.2 %
• Output Accuracy = Output Range Span (20 mA) ÷ Output Span Setting (16 mA) × 0.06 % = 0.075 %
Overall Accuracy = 0.2 + 0.075 = ±0.275 %
STANDARDS & APPROVALS
EU conformity
:
EMC Directive
EMI EN 61000-6-4
EMS EN 61000-6-2
Low Voltage Directive
EN 61010-1
Installation Category II
Pollution Degree 2
Input or output to power: Reinforced insulation (300 V)
RoHS Directive
EXTERNAL VIEW
Refer to the operating manual (EM-5984-B) for detailed procedures.
TERMINAL ASSIGNMENTS
* Base does not come with the unit. Please order separately.
PARAMETER LIST
EXTERNAL DIMENSIONS & TERMINAL ASSIGNMENTS unit: mm [inch]
MOUNTING REQUIREMENTS unit: mm [inch]
* Mounting requirements for base.
SCHEMATIC CIRCUITRY & CONNECTION DIAGRAM